Related Experiment Video
Updated: Apr 5, 2026

Diffusion Tensor Magnetic Resonance Imaging in Chronic Spinal Cord Compression
Published on: May 7, 2019
Diffusion-prepared neurography of the brachial plexus with a large field-of-view at 3T
Jos Oudeman1, Bram F Coolen1, Valentina Mazzoli1,2,3
1Department of Radiology, Academic Medical Center, Amsterdam, The Netherlands.
Purpose:
To study diffusion-prepared neurography optimized for a large field-of-view (FOV) to include the neck and both shoulders. In a large FOV poor homogeneity of the magnetic field (B0 ) often leads to poor image quality and possibly to poor diagnostic accuracy. The aim was therefore to find an optimal (combination of) shimming method(s) for diffusion-prepared neurography in a large FOV.
Materials And Methods:
A 3D diffusion-prepared sequence with a large FOV was tested with and without the use of a susceptibility-matched pillow combined with image-based (IB) or standard shimming in six healthy volunteers on a 3T system. B0 , B1 , signal to noise ratio (SNR), and contrast to noise ratio (CNR) were compared between all protocols. Additionally, nerve visibility, fat suppression, artifacts, and overall image quality were ordinally (5-point scale) assessed by two readers. Furthermore, correlations between B0 and B1 (offset and variation) and SNR, CNR, and image quality were explored.
Results:
The use of the susceptibility-matched pillow led to a 43% reduction of B0 variation over the brachial plexus compared to the situation without a pillow (P < 0.05). The combination of the pillow with IB-shimming and the optimized diffusion-prepared sequence resulted in good nerve visibility, good fat suppression, no artifacts that would hinder clinical diagnosis, and a good overall quality (median scores ≥4). Reducing B0 variation was associated with SNR, CNR, and the above-mentioned scored features (P < 0.05).
Conclusion:
The use of a susceptibility-matched pillow in combination with IB-shimming enables robust and high-quality neurography of the complete brachial plexus.