Highly sensitive and selective uric acid biosensor based on a three-dimensional graphene foam/indium tin oxide glass
Hong Yan Yue1, Hong Zhang1, Jing Chang1
1School of Materials Science and Engineering, Harbin University of Science and Technology, Harbin 150040, PR China.
Abstract:
A three-dimensional (3D) continuous and interconnected network graphene foam (GF) was synthesized by chemical vapor deposition using nickel foam as a template. The morphologies of the GF were observed by scanning electron microscopy. X-ray diffraction and Raman spectroscopy were used to investigate the structure of GF. The graphene with few layers and defect free was closely coated on the backbone of the 3D nickel foam. After etching nickel, the GF was transferred onto indium tin oxide (ITO) glass, which acted as an electrode to detect uric acid using cyclic voltammetry (CV) and differential pulse voltammetry (DPV). The GF/ITO electrode showed a high sensitivity for the detection of uric acid: approximately 9.44 mA mM(-1) in the range of 25 nM-0.1 μM and 1.85 mA mM(-1) in the range of 0.1-60 μM. The limit of detection of GF/ITO electrode for uric acid is 3 nM. The GF/ITO electrode also showed a high selectivity for the detection of uric acid in the presence of ascorbic acid. This electrode will have a wide range of potential application prospects in electrochemical detection.