CD4+ T-Lymphocyte Capture Using a Disposable Microfluidic Chip for HIV
Published on: October 1, 2007
Electrochemical magneto-actuated biosensor for CD4 count in AIDS diagnosis and monitoring
S Carinelli1, C Xufré Ballesteros2, M Martí2
1Grup de Sensors i Biosensors, Departament de Química, Universitat Autònoma de Barcelona, Spain.
Abstract:
The counting of CD4(+) T lymphocytes is a clinical parameter used for AIDS diagnosis and follow-up. As this disease is particularly prevalent in developing countries, simple and affordable CD4 cell counting methods are urgently needed in resource-limited settings. This paper describes an electrochemical magneto-actuated biosensor for CD4 count in whole blood. The CD4(+) T lymphocytes were isolated, preconcentrated and labeled from 100 μL of whole blood by immunomagnetic separation with magnetic particles modified with antiCD3 antibodies. The captured cells were labeled with a biotinylated antiCD4 antibody, followed by the reaction with the electrochemical reporter streptavidin-peroxidase conjugate. The limit of detection for the CD4 counting magneto-actuated biosensor in whole blood was as low as 44 cells μL(-1) while the logistic range was found to be from 89 to 912 cells μL(-1), which spans the whole medical interest range for CD4 counts in AIDS patients. The electrochemical detection together with the immunomagnetic separation confers high sensitivity, resulting in a rapid, inexpensive, robust, user-friendly method for CD4 counting. This approach is a promising alternative for the costly standard flow cytometry and suitable as diagnostic tool at decentralized practitioner sites in low resource settings, especially in less developed countries.