A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Sinus rhythm R-wave amplitude as a predictor of ventricular fibrillation undersensing in patients with implantable
Linda L Ruetz1, Jodi L Koehler1, Mark L Brown1
1Medtronic, Inc., Minneapolis, Minnesota.
Background:
Ventricular fibrillation (VF) is induced during implantable cardioverter-defibrillator (ICD) implantation to ensure that the ICD will sense, detect, and defibrillate VF. ICD implant guidelines state that the amplitude of the sinus rhythm R wave recorded from the ventricular electrogram should have amplitude ≥5 mV. No study has tested the relationship between sinus rhythm R-wave amplitude and VF sensing using modern, transvenous sensing electrodes.
Objective:
The goal of this study was to determine whether there is a sinus rhythm R-wave amplitude cutoff that can be used to determine which patients are not at risk of VF undersensing.
Methods:
A retrospective analysis of induced and spontaneous VF episodes from 2 clinical trials with 2022 patients was performed. Episodes with undersensing during the initial detection of VF were identified, and the distribution of sinus rhythm R-wave amplitudes for patients with and without VF undersensing was analyzed.
Results:
Only 3% of analyzed induced VF episodes were considered to have VF undersensing, and none had clinically significant detection delays. There was no correlation between device-measured, rectified sinus rhythm R-wave amplitude and VF undersensing at the time of implantation or during follow-up, although <4% of patients had sinus rhythm R-waves with amplitude <3 mV.
Conclusion:
We analyzed true bipolar sensing of induced VF or spontaneous ventricular tachycardia/VF detected in the ICD VF zone. Sensing of VF was so reliable that clinically significant undersensing did not occur. Our findings do not support any recommended minimum sinus rhythm R wave to ensure reliable sensing of VF or the necessity of inducing VF to verify sensing for rectified sinus rhythm R-waves with amplitude ≥3 mV.
More Related Videos
Related Concept Videos
ECG Interpretation of Arrhythmias I: Sinus Arrhythmias
Types of Arrhythmias
Sinus Node Arrhythmias
Sinus Bradycardia: Originating from the sinoatrial (SA) node, sinus bradycardia involves slower impulses, resulting in a heart rate of less than 60 beats per minute (bpm). Causes include sleep, vagal stimulation, beta-blockers, hypothyroidism,...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias
Dysrhythmias V: Evaluating Dysrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...