Published on: December 14, 2011
Reliability of a new method for measuring coronal trunk imbalance, the axis-line-angle technique
Rui-Fang Zhang1, Kun Liu1, Xue Wang1
1Department of Radiology, The Second Affiliated Hospital of Wenzhou Medical University, 109 Xueyuanxi Rd, 325027 Wenzhou, China.
Background Context:
Accurate determination of the extent of trunk imbalance in the coronal plane plays a key role in an evaluation of patients with trunk imbalance, such as patients with adolescent idiopathic scoliosis. An established, widely used practice in evaluating trunk imbalance is to drop a plumb line from the C7 vertebra to a key reference axis, the central sacral vertical line (CSVL) in full-spine standing anterioposterior radiographs, and measuring the distance between them, the C7-CSVL. However, measuring the CSVL is subject to intraobserver differences, is error-prone, and is of poor reliability. Therefore, the development of a different way to measure trunk imbalance is needed.
Purpose:
This study aimed to describe a new method to measure coronal trunk imbalance, the axis-line-angle technique (ALAT), which measures the angle at the intersection between the C7 plumb line and an axis line drawn from the vertebral centroid of the C7 to the middle of the superior border of the symphysis pubis, and to compare the reliability of the ALAT with that of the C7-CSVL.
Study Design/Setting:
A prospective study at a university hospital was used.
Patient Sample:
The patient sample consisted of sixty-nine consecutively enrolled men and women patients, aged 10-18 years, who had trunk imbalance defined as C7-CSVL longer than 20 mm on computed full-spine standing anterioposterior radiographs.
Outcome Measures:
Data were analyzed to determine the correlation between C7-CSVL and ALAT measurements and to determine intraobserver and interobserver reliabilities.
Methods:
Using a picture archiving and communication system, three radiologists independently evaluated trunk imbalance on the 69 computed radiographs by measuring the C7-CSVL and by measuring the angle determined by the ALAT. Data were analyzed to determine the correlations between the two measures of trunk imbalance, and to determine intraobserver and interobserver reliabilities of each of them.
Results:
Overall results from the measurements by the C7-CSVL and the ALAT were significantly moderately correlated. Intraobserver assessments by measuring the C7-CSVL and by doing the ALAT failed to find any significant differences between the findings from the first and second assessments by the same radiologist. Interobserver assessments significantly differed between radiologists 1 and 2 for the first assessment measuring the C7-CSVL, and between radiologists 2 and 3 for the second assessment measuring the C7-CSVL. Interobserver assessments by doing the ALAT failed to find any significant differences among the three radiologists for either of the two assessments.
Conclusions:
Our results indicated that using the ALAT, which is simple and convenient, is of great value in measuring trunk imbalance. For measuring trunk imbalance, the ALAT has essential advantages compared with measuring the C7-CSVL. We encourage spine surgeons to consider using the ALAT in evaluating trunk imbalance.
More Related Videos
10:07Measurement of Dynamic Scapular Kinematics Using an Acromion Marker Cluster to Minimize Skin Movement Artifact
Published on: February 10, 2015
08:40Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Related Concept Videos
Adjusting a Traverse
Unsymmetric Bending - Angle of Neutral Axis
When a bending moment is applied at an angle θ concerning the vertical axis of a symmetrical member, it can be resolved into components along the member's principal...