Objectives:
Fever screening systems, such as Infrared Thermal Detection Systems (ITDS), have been used for rapid identification of potential cases during respiratory disease outbreaks for public health management. ITDS detect a difference between the subject and ambient temperature, making deployment in hot climates more challenging. This study, conducted in Singapore, a tropical city, evaluates the accuracy of three different ITDS for fever detection compared with traditional oral thermometry and self-reporting in a clinical setting.
Study Design:
This study is a prospective operational evaluation conducted in the Singapore military on all personnel seeking medical care at a high-volume primary healthcare centre over a one week period in February 2014.
Methods:
Three ITDS, the STE Infrared Fever Screening System (IFSS), the Omnisense Sentry MKIII and the handheld Quick Shot Infrared Thermoscope HT-F03B, were evaluated. Temperature measurements were taken outside the healthcare centre, under a sheltered walkway and compared to oral temperature. Subjects were asked if they had fever.
Results:
There were 430 subjects screened, of whom 34 participants (7.9%) had confirmed fever, determined by oral thermometer measurement. The handheld infrared thermoscope had a very low sensitivity (29.4%), but a high specificity (96.8%). The STE ITDS had a moderate sensitivity (44.1%), but a very high specificity (99.1%). Self-reported fevers showed good sensitivity (88.2%) and specificity (93.9%). The sensitivity of the Omnisense ITDS (89.7%) was the highest among the three methods with good specificity (92.0%).
Conclusion:
The new generation Omnisense ITDS displayed a relatively high sensitivity and specificity for fever. Though it has a lower sensitivity, the old generation STE ITDS system showed a very high specificity. Self-reporting of fever was reliable. The handheld thermograph should not be used as a fever-screening tool under tropical conditions.
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.