Dose reduction with iterative reconstruction in multi-detector CT: What is the impact on deformation of circular
J Greffier1, F Macri1, A Larbi1
1EA 2415, Department of Radiology, Nîmes University Hospital, boulevard Professeur-Robert-Debré, 30029 Nîmes cedex, France.
Objectives:
To evaluate the distortion of circular structures induced by the increased image noise related to dose reduction and to assess the effect of iterative reconstruction (IR).
Methods:
MDCT acquisitions were performed with 120 kVp for 200/100/60/40/20mAs with 100%/50%/30%/20%/10% of dose. Raw data were reconstructed by filtered back projection (FBP) and with two IR strengths. Image quality indices referred to water and acrylic were measured on a quality image phantom. Areas, perimeters, circularity were measured on the circular inserts of 4.8, 7.9 and 11.1mm on a morphological phantom.
Results:
Dose reduction resulted in increased image noise and in decreased signal to noise ratio and contrast to noise ratio. IR improved these indices for the same dose without affecting the signal (number CT) and spatial resolution (modulation transfer function). The values of area, perimeter and circularity were altered compared to the actual value and the inserts were visually deformed with the dose reduction. IR improved these three parameters. Image quality indices, areas, perimeters and circularity of inserts were similar between the acquisition at 100% of the dose in FBP, 50% in strength-3 and 30% in strength-5 with different curves of noise power spectrum.
Conclusion:
IR associated to 70% of dose reduction modifies the images smooth (NPS) but maintains adequate image quality indices without causing distortions of circular structures.
Related Concept Videos
Deformation in a Circular Shaft
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Plastic Deformation in Circular Shafts