Dual-Tuned 31P-1H Dual-Row Loop/Dipole 32-Element Transceiver Array for Human Brain Spectroscopy at 9.4T
Related Experiment Video
Updated: Oct 31, 2026

Mapping the After-effects of Theta Burst Stimulation on the Human Auditory Cortex with Functional Imaging
Published on: September 12, 2012
G A Solomakha1, R Pohmann1, F Glang1,2
1High-Field MR Center, Max Planck Institute for Biological Cybernetics, Tuebingen, Germany.
Purpose:
The goal of this work is to develop and evaluate a single-layer tight-fit 32-element dual-tuned (DT) loop/dipole transceiver (TxRx) array for human brain 31P MRS(I) at 9.4T, achieving full-brain coverage and reasonable transmit and receive performance at both frequencies.
Methods:
First, we developed numerical models of dual-row TxRx arrays for 31P (loop array) and 1H (coaxial-end folded-end dipole array) frequencies at 9.4T. Next, multi-tissue voxel models were used to simulate Tx-performance of the arrays and define the optimal CP-mode excitation. Following this, the proposed array's performance was evaluated by MR measurements both on a phantom and a healthy volunteer. Finally, we compared the proposed array to the previously reported DT single-row loop-based TxRx array.
Results:
The developed 32-element DT TxRx array demonstrated full-brain (including the cerebellum and brain stem) imaging capabilities, reasonable SNR, and transmit performance at both frequencies at 9.4T.
Conclusion:
As a proof of concept, we developed the 32-element DT UHF tight-fit TxRx human head array coil for 31P MRS(I) with good 1H performance using a combination of loops and dipoles. The proposed array design could also be adapted to higher fields, that is, 10.5T, 11.7T, and 14T.