X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
SCATTERED X-RAY ENERGY DATA FROM INTEGRATED MULTI-FILTER PERSONAL DOSEMETERS WORN BY INTERVENTIONAL RADIOLOGY STAFF
1Department of Radiology, Japan Community Healthcare Organization Hokkaido Hospital, 1-8-3-18 Nakanoshima, Toyohira, Sapporo, Hokkaido 062-8618, Japan mori-hiroshige@hokkaido.jcho.go.jp.
Abstract:
The accuracy and usability of scattered X-ray energy data obtained from multi-filter dosemeters were investigated by comparing the data with results obtained via Monte Carlo simulation. The energy data, which were read from integrated personal dosemeters used by interventional radiology (IR) staff in individual monitoring, were plotted against relative frequency and statistically assessed. The effective energy obtained from the multi-filter dosemeters was inversely proportional to detector size and lower than the results of the Monte Carlo simulation. All distributions of scattered X-ray energy to which the IR staff were exposed had only one peak value, which corresponded with the effective energy measured using a water phantom. For abdominal IR, there was a significant difference in the distributions between IR staff (p < 0.01). The results suggest that rectifying systematic errors resulting from oblique incidence of X-rays against the filters would make the energy data usable for optimisation of radiological protection.
More Related Videos
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT