Lead screening in DBS by solid sampling high-resolution continuum source graphite furnace atomic absorption
Luis Rello1, Maite Aramendía2,3, Miguel A Belarra3
1Department of Clinical Biochemistry, Hospital Universitario "Miguel Servet", Paseo Isabel La Católica 1-3, 50009, Zaragoza, Spain.
Background:
DBS have become a clinical specimen especially adequate for establishing home-based collection protocols. In this work, high-resolution continuum source graphite furnace atomic absorption spectrometry is evaluated for the direct monitoring of Pb in DBS, both as a quantitative tool and a screening method.
Methodology:
The development of the screening model is based on the establishment of the unreliability region around the threshold limits, 100 or 50 μg l(-1). More than 500 samples were analyzed to validate the model.
Conclusion:
The screening method demonstrated high sensitivity (the rate of true positives detected was always higher than 95%), an excellent LOD (1 µg l(-1)) and high throughput (10 min per sample).
More Related Videos
11:33All-electronic Nanosecond-resolved Scanning Tunneling Microscopy: Facilitating the Investigation of Single Dopant Charge Dynamics
Published on: January 19, 2018
13:09Assessment of Boron Doped Diamond Electrode Quality and Application to In Situ Modification of Local pH by Water Electrolysis
Published on: January 6, 2016
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Overview
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Emission Spectroscopy: Instrumentation