Feasibility study on signal separation for spontaneous alpha decay in LaBr3: Ce scintillator by signal peak-to-charge
1Department of Medical Physics and Engineering, Hokkaido University Graduate School of Medicine, Kita-15 Nishi-7, Kita-ku, Sapporo-shi, Hokkaido, Japan.
Abstract:
A novel analysis method named peak-to-charge ratio (V(p)/Q(total)) discrimination, aiming at background rejection especially for alpha decay self-activity in LaBr3: Ce scintillators has been developed. This method is based on a waveform analysis using the peak-to-charge ratio in the output waveform of a photomultiplier tube. The discrimination of alpha-induced events was achieved by using a threshold function based on the error propagation of the V(p)/Q(total) value. The accidental rejection ratio of gamma-induced events was evaluated to be 0.17%. Furthermore, a total absorption peak spectrum processed with the V(p)/Q(total) discrimination method for (68)Ga 1.883 MeV gamma rays, where the energy was overlapped with background alpha events, reproduced exactly the same result as that of the background subtraction method. The difference in measured peak counts of both methods was 0.716%, and the statistical error in the V(p)/Q(total) discrimination method and background subtraction was 4.81% and 8.70%, respectively. Thus a higher-accuracy measurement could be achieved using the V(p)/Q(total) discrimination method. The present study demonstrates that the V(p)/Q(total) discrimination method is a promising method for background rejection of the spontaneous alpha decay in LaBr3:Ce scintillators.
Related Concept Videos
¹³C NMR: ¹H–¹³C Decoupling
A broadband decoupling technique is used to simplify these complex, sometimes overlapping, signals. Broadband decoupling relies on a...
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Overview