Inappropriate mode switching clarified by using a chest radiograph.
Brian Marino1, Abhishek Jaiswal2, Seth Goldbarg3
1Division of Cardiology, Department of Internal Medicine, New York Hospital Queens/Cornell University Medical College, WA 200, 56-45 Main Street, Flushing, NY 11355, USA.
Pacemaker lead-lead interaction caused artifacts and oversensing, leading to inappropriate mode switching in a patient with atrial fibrillation and atrioventricular node disease. This highlights a potential cause of pacemaker malfunction during procedures.
Area of Science:
- Cardiology
- Biomedical Engineering
- Electrophysiology
Background:
- Dual-chamber pacemakers are used for bradyarrhythmias like atrioventricular node disease.
- Paroxysmal atrial fibrillation can complicate pacemaker management.
- Percutaneous coronary interventions (PCI) require careful monitoring of implanted devices.
Purpose of the Study:
- To investigate the cause of abnormal pacing episodes and inappropriate mode switching during PCI.
- To identify the source of nonphysiologic recordings in a pacemaker patient.
- To evaluate potential lead-related issues causing pacemaker malfunction.
Main Methods:
- Pacemaker interrogation to assess device function and pacing parameters.
- Analysis of representative electrograms to identify signal characteristics.
- Chest radiography to evaluate lead positioning and integrity.
- Exclusion of intrinsic pacemaker malfunction through diagnostic testing.
Main Results:
- Frequent short-duration mode switching episodes were detected.
- High-frequency, nonphysiologic recordings were predominantly observed in the atrial lead.
- Excessive atrial and ventricular lead slack was identified in the right ventricular inflow.
- Intrinsic pacemaker malfunction was ruled out as the cause.
Conclusions:
- Lead-lead interaction is suspected as the cause of artifacts and oversensing.
- This interaction led to frequent, inappropriate mode switching episodes.
- Lead management and positioning are critical for preventing pacemaker malfunction, especially during invasive procedures.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Pneumothorax-II
Clinical Manifestations:
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...