Post Column Derivatization Using Reaction Flow High Performance Liquid Chromatography Columns
Published on: April 26, 2016
Post column derivatisation analyses review. Is post-column derivatisation incompatible with modern HPLC columns?
Andrew Jones1, Sercan Pravadali-Cekic1, Gary R Dennis1
1Australian Centre for Research on Separation Sciences (ACROSS), School of Science and Health, University of Western Sydney (Parramatta), NSW 2150, Australia.
Post-column derivatization (PCD) methods often use large reaction coils, causing peak broadening and reduced separation efficiency. This incompatibility hinders the adoption of modern ultra-high performance liquid chromatography (UHPLC) techniques.
Area of Science:
- Analytical Chemistry
- Chromatography
Background:
- Post-column derivatization (PCD) coupled with High-Performance Liquid Chromatography (HPLC) and Ultra-High-Performance Liquid Chromatography (UHPLC) is a key analytical technique.
- A significant drawback of PCD is the increased post-column dead volume from reaction coils, leading to peak broadening and reduced separation efficiency.
Purpose of the Study:
- To review the compatibility of existing PCD methods with modern UHPLC systems.
- To analyze system parameters in published PCD methods, focusing on column dimensions, stationary phase, particle size, and reaction loop geometry.
Main Methods:
- A systematic review of 87 PCD methods published between 2009 and 2014.
- Analysis of column dimensions (length, internal diameter, particle size) and reaction loop volumes.
- Experimental evaluation of dead volume effects on column performance (resolution, separation efficiency).
Main Results:
- Most reviewed PCD methods utilized conventional HPLC columns (e.g., 250 × 4.6 mm i.d. with 5 μm particles) rather than UHPLC-compatible columns.
- The majority of methods employed reaction loops exceeding 500 μL.
- Experimental data demonstrated a severe loss of performance with reaction loops larger than 100 μL.
Conclusions:
- Practitioners of PCD methods predominantly use column formats incompatible with UHPLC.
- The dead volume introduced by typical PCD reaction coils significantly compromises the efficiency gains offered by modern HPLC and UHPLC technologies.
- There is a fundamental incompatibility between conventional PCD setups and the requirements of high-efficiency UHPLC separations.
Related Concept Videos
Principles Of Column Chromatography
Column Efficiency: Plate Theory
A higher number of theoretical plates signifies better column efficiency and improved separation capabilities. Plate height affects bandwidth and separation quality; it is inversely...
First Derivative Test: Problem Solving
Types Of Column Chromatography
Gel Filtration Chromatography
When the...
Relative Reactivity of Carboxylic Acid Derivatives
A key factor in assessing the reactivity of the acid derivatives is the basicity of the substituent or the leaving group. The lower the basicity of the leaving group, the higher the reactivity of the derivative. The basicity of the leaving group follows this order:
Halide ions < Acyloxy ions < Alkoxy ions < Amine ions
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)