Agreement Between Doppler and Invasive Blood Pressure Monitoring in Anesthetized Dogs Weighing <5 kg
Martin J Kennedy1, Michele Barletta
1From the Department of Anesthesia and Pain Management, University of Wisconsin Veterinary Medical Teaching Hospital, Madison, WI (M.K.); and Anesthesia Section, University of Minnesota, Veterinary Medical Center, Saint Paul, MN (M.B.).
Abstract:
The objectives of this study were to determine if Doppler (DOP) blood pressure measurements more closely estimate either invasive systolic or invasive mean arterial blood pressures (ISAP or IMAP, respectively) in small dogs under general anesthesia and to assess the ability of DOP to detect anesthesia-related hypotension in small dogs. Blood pressure measurements (n = 203) were obtained from 10 client-owned dogs. DOP, ISAP, and IMAP were recorded simultaneously, and the data were categorized into two groups: hypotensive (ISAP <90 mm Hg) and normotensive (ISAP ≥90 mm Hg and ≤160 mm Hg). DOP overestimated ISAP and IMAP in both the normotensive and hypotensive groups. The DOP was highly specific (97%) but poorly sensitive (56%) for detecting hypotension. The smallest bias was achieved when using DOP as an estimate of systolic arterial blood pressure in both normotensive and hypotensive dogs, suggesting that DOP measures systolic arterial blood pressure in dogs <5 kg. For dogs with hypotension, DOP met all of the performance criteria for noninvasive blood pressure monitors recommended by the American College of Veterinary Internal Medicine. DOP is an acceptably accurate and highly specific means of detecting hypotension in small dogs under general anesthesia.
More Related Videos
08:12Invasive Hemodynamic Monitoring of Aortic and Pulmonary Artery Hemodynamics in a Large Animal Model of ARDS
Published on: November 26, 2018
08:51Multiple Intravenous Bolus Dosing and Invasive Hemodynamic Assessment in a Hypoxia-Induced Mouse Pulmonary Artery Hypertension Model
Published on: November 11, 2022
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure