Related Experiment Video
Updated: Apr 4, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Parallelized patient-specific quality assurance for high-dose-rate image-guided brachytherapy in an integrated
Taeho Kim1, Timothy N Showalter1, W Tyler Watkins1
1Department of Radiation Oncology, University of Virginia Health System, Charlottesville, VA.
Purpose:
To describe a parallelized patient-specific quality assurance (QA) program designed to ensure safety and quality in image-guided high-dose-rate brachytherapy in an integrated computed tomography (CT)-on-rails brachytherapy suite.
Materials And Methods:
A patient-specific QA program has been modified for the image-guided brachytherapy (IGBT) program in an integrated CT-on-rails brachytherapy suite. In the modification of the QA procedures of Task Group-59, the additional patient-specific QA procedures are included to improve rapid IGBT workflow with applicator placement, imaging, planning, treatment, and applicator removal taking place in one room.
Results:
The IGBT workflow is partitioned into two groups of tasks that can be performed in parallel by two or more staff members. One of the unique components of our implemented workflow is that groups work together to perform QA steps in parallel and in series during treatment planning and contouring. Coordinating efforts in this systematic way enable rapid and safe brachytherapy treatment while incorporating 3-dimensional anatomic variations between treatment days.
Conclusions:
Implementation of these patient-specific QA procedures in an integrated CT-on-rails brachytherapy suite ensures confidence that a rapid workflow IGBT program can be implemented without sacrificing patient safety or quality and deliver highly-conformal dose to target volumes. These patient-specific QA components may be adapted to other IGBT environments that seek to provide rapid workflow while ensuring quality.