Sevoflurane Output in the Isoflurane/Halothane Diamedica Draw-over Vaporiser
1Department of Anaesthesia, National University Hospital, Singapore.
Introduction:
Draw-over anaesthesia remains an attractive option for conduct of anaesthesia in austere conditions. The Diamedica Draw-over Vaporiser (DDV) is a modern draw-over vaporiser and has separate models for isoflurane/halothane and sevoflurane.
Materials And Methods:
A laboratory study was done to measure sevoflurane output in an isoflurane/ halothane DDV. We did 3 series of experiments with the isoflurane/halothane DDV. We measured anaesthetic agent output in both push-over and draw-over setups, and at minute ventilation of 6 L/min and 3 L/min. Series 1 experiment was done with isoflurane in the DDV at ambient temperature of 20°C. Series 2 experiment was done with sevoflurane in the DDV at ambient temperature of 20°C. Series 3 experiment was done with sevoflurane in the DDV and with the DDV placed in a water bath of 40°C.
Results:
The sevoflurane output was found to be two-thirds of the isoflurane/ halothane DDV dial setting at ambient temperature of 20°C. With the DDV in a 40°C water bath, the sevoflurane output was found to be about the isoflurane/ halothane DDV dial settings.
Conclusion:
In our experiment, we show that it is possible to use sevoflurane in an isoflurane/halothane DDV.
More Related Videos
Related Concept Videos
Inhalational Anesthetics: Overview
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Vaporization
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...