Effects of Needle and Catheter Size on Commercially Available Ultrasound Contrast Agents
John R Eisenbrey1, Annemarie Daecher2, Michael R Kramer2
1Department of Radiology, Thomas Jefferson University, Philadelphia, Pennsylvania USA (J.R.E., A.D., M.R.K., F.F.); Villanova University, Villanova, Pennsylvania USA (A.D.); and School of Medicine, Temple University, Philadelphia, Pennsylvania USA (M.R.K.). john.eisenbrey@jefferson.edu.
Objectives:
To investigate effects of needle and catheter size on in vitro ultrasound contrast agent (UCA) enhancement and concentrations using 4 commercially available UCAs.
Methods:
Definity (Lantheus Medical Imaging, North Billerica, MA), Optison (GE Healthcare, Princeton, NJ), SonoVue (Bracco SA, Geneva, Switzerland), and Sonazoid (GE Healthcare, Oslo, Norway) were investigated. The UCA was injected via a 1-mL syringe (BD, Franklin Lakes, NJ) into a 3-way stopcock (Smith Medical, Dublin, OH) and flushed with 10 mL of saline through an 18-cm infusion extension tube connected to either a 16-, 18-, 20-, 22-, or 24-gauge catheter (BD) or an 18-, 20-, 21-, or 25-gauge needle (BD). In vitro enhancement was determined in a flow phantom (ATS Laboratories, Bridgeport, CT), and microbubble concentrations were determined using an LSRII flow cytometer (BD Biosciences, San Jose, CA).
Results:
Significant decreases in enhancement and microbubble concentrations were observed for all 4 UCAs (P < .001) when administration was performed through a 25-gauge needle. No statistically significant differences in enhancement or concentrations were observed between all catheter sizes and 18- to 21-gauge needles for SonoVue and Sonazoid. Definity and Optison administration through a 24-gauge catheter resulted in a significant loss of enhancement (P < .02), although these differences were not significant on flow cytometry.
Conclusions:
Administration of commercial UCAs in a clinical scenario is possible with catheters or needles smaller than 20 gauge, although the minimal allowable size appears to be UCA specific.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies VII: Vascular Imaging
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...