Long-term strategies support autonomy in radiation safety in invasive cardiology
Eberhard Kuon1, Stephan B Felix2, Kerstin Weitmann3
1Department of Cardiology, Klinik Fraenkische Schweiz, Ebermannstadt, Germany.
Background And Purpose:
Despite comprehensive radiation safety programs, radiation exposure in invasive cardiology remains considerable. According to the 2013 German Registry, median in-hospital dose area products (DAP) amount to 19.8Gycm(2) for invasive coronary angiography (CA). We analyzed long-term radiation-reducing strategies for an experienced interventionalist from 1997 to 2012, for the target intervention of CA.
Methods:
Among representative cohorts, we evaluated iterative alterations in collimation, time on beam, pulse rates, detector entrance doses, and angulations on the basis of DAP, radiographic DAP(R) and fluoroscopic DAP(F), the respective times on beam, and the number of frames and runs.
Results:
Patients' median overall DAP decreased from 33.8Gycm(2) at baseline to 2.4 and 0.6Gycm(2) for CA in conventional (C) and electrocardiogram-gated (E) modes - one diastolic radiographic frame per heartbeat at 77% of the RR interval. Further median dose parameters for CA at baseline and finally in C/E mode were as follows: effective dose (6.76-0.48/0.13mSv), radiography time (43.8-12.9/21.7s), frames (548-105/25), frames/run (41.3-14.4/3.4), DAP(R)/frame (42.6-16.6/12.6mGycm(2)), DAP(R)/s (532-130/13.8mGycm(2)/s), fluoroscopy time (195-120/119s), DAP(F)/pulse (2.0-1.1/0.8mGycm(2)), and DAP(F)/s (48.9-4.4/3.1mGycm(2)/s).
Conclusions:
Our data highlight the efficacy of various radiation-reducing strategies by autonomous control and iterative training in radiation safety toward submillisievert levels for CA, and define realizable benchmarks for comparison with the performance data of any individual.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Myocarditis IV: Nursing Management
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cardiac Catheterization IV: Nursing Management
Cardiomyopathy V: Interprofessional Care
Endocarditis IV: Nursing Management