Related Experiment Video
Updated: Apr 3, 2026

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Calculation of Organ Doses for a Large Number of Patients Undergoing CT Examinations
Amir Bahadori1, Diana Miglioretti2, Randell Kruger3
11 Space Radiation Analysis Group, National Aeronautics and Space Administration, Johnson Space Center, Houston, TX.
Objective:
The objective of our study was to develop an automated calculation method to provide organ dose assessment for a large cohort of pediatric and adult patients undergoing CT examinations.
Materials And Methods:
We adopted two dose libraries that were previously published: the volume CT dose index-normalized organ dose library and the tube current-exposure time product (100 mAs)-normalized weighted CT dose index library. We developed an algorithm to calculate organ doses using the two dose libraries and the CT parameters available from DICOM data. We calculated organ doses for pediatric (n = 2499) and adult (n = 2043) CT examinations randomly selected from four health care systems in the United States and compared the adult organ doses with the values calculated from the ImPACT calculator.
Results:
The median brain dose was 20 mGy (pediatric) and 24 mGy (adult), and the brain dose was greater than 40 mGy for 11% (pediatric) and 18% (adult) of the head CT studies. Both the National Cancer Institute (NCI) and ImPACT methods provided similar organ doses (median discrepancy < 20%) for all organs except the organs located close to the scanning boundaries. The visual comparisons of scanning coverage and phantom anatomies revealed that the NCI method, which is based on realistic computational phantoms, provides more accurate organ doses than the ImPACT method.
Conclusion:
The automated organ dose calculation method developed in this study reduces the time needed to calculate doses for a large number of patients. We have successfully used this method for a variety of CT-related studies including retrospective epidemiologic studies and CT dose trend analysis studies.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET