Atrioventricular and ventricular-to-ventricular programming in patients with cardiac resynchronization therapy:
Benjamin A Steinberg1,2, Scott Wehrenberg3, Kevin P Jackson4
1Electrophysiology Section, Duke University Medical Center, PO Box 17969, Durham, NC, 27715, USA. benjamin.steinberg@duke.edu.
Purpose:
Cardiac resynchronization therapy (CRT) improves outcomes in patients with heart failure, yet response rates are variable. We sought to determine whether physician-specified CRT programming was associated with improved outcomes.
Methods:
Using data from the ALTITUDE remote follow-up cohort, we examined sensed atrioventricular (AV) and ventricular-to-ventricular (VV) programming and their associated outcomes in patients with de novo CRT from 2009-2010. Outcomes included arrhythmia burden, left ventricular (LV) pacing, and all-cause mortality at 4 years.
Results:
We identified 5709 patients with de novo CRT devices; at the time of implant, 34% (n = 1959) had entirely nominal settings programmed, 40% (n = 2294) had only AV timing adjusted, 11% (n = 604) had only VV timing adjusted, and 15% (n = 852) had both AV and VV adjusted from nominal programming. Suboptimal LV pacing (<95%) during follow-up was similar across groups; however, the proportion with atrial fibrillation (AF) burden >5% was lowest in the AV-only adjusted group (17.9%) and highest in the nominal (27.7%) and VV-only adjusted (28.3%) groups. Adjusted all-cause mortality was significantly higher among patients with non-nominal AV delay >120 vs. <120 ms (adjusted heart rate (HR) 1.28, p = 0.008) but similar when using the 180-ms cutoff (adjusted HR 1.13 for >180 vs. ≤180 ms, p = 0.4).
Conclusions:
Nominal settings for de novo CRT implants are frequently altered, most commonly the AV delay. There is wide variability in reprogramming. Patients with nominal or AV-only adjustments appear to have favorable pacing and arrhythmia outcomes. Sensed AV delays less than 120 ms are associated with improved survival.
More Related Videos
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Dysrhythmias VI: Management of Dysrhythmias
Electrophysiology of Normal Cardiac Rhythm
Cardiopulmonary Resuscitation III: AED Use