Performance of mobile digital X-ray fluoroscopy using a novel flat panel detector for intraoperative use
Chang-Won Jeong1, Jong-Hyun Ryu1, Su-Chong Joo2
1Imaging Science Research Center, Wonkwang University School of Medicine, Iksan, Jeonbuk, Korea.
Background:
Technologies employing digital X-ray devices are developed for mobile settings.
Objective:
To develop a mobile digital X-ray fluoroscopy (MDF) for intraoperative guidance, using a novel flat panel detector to focus on diagnostics in outpatient clinics, operating and emergency rooms.
Methods:
An MDF for small-scale field diagnostics was configured using an X-ray source and a novel flat panel detector. The imager enabled frame rates reaching 30 fps in full resolution fluoroscopy with maximal running time of 5 minutes. Signal-to-noise (SNR), contrast-to-noise (CNR), and spatial resolution were analyzed. Stray radiation, exposure radiation dose, and effective absorption dose were measured for patients.
Results:
The system was suitable for small-scale field diagnostics. SNR and CNR were 62.4 and 72.0. Performance at 10% of MTF was 9.6 lp/mm (53 μ m) in the no binned mode. Stray radiation at 100 cm and 150 cm from the source was below 0.2 μ Gy and 0.1 μ Gy. Exposure radiation in radiography and fluoroscopy (5 min) was 10.2 μ Gy and 82.6 mGy. The effective doses during 5-min-long fluoroscopy were 0.26 mSv (wrist), 0.28 mSv (elbow), 0.29 mSv (ankle), and 0.31 mSv (knee).
Conclusions:
The proposed MDF is suitable for imaging in operating rooms.