Research on an anti-motion interference algorithm of blood oxygen saturation based on AC and DC analysis
Background:
There is an urgent need for blood oxygen saturation (SpO2) tests when participants are ambulatory, as in daily activity monitoring, sleep monitoring, or even athletes' cardiovascular function tests. In such situations, measuring equipment needs to be wearable. This restricts the processor volume, and the corresponding algorithm should be microprocessor compatible.
Objective:
This article proposes an anti-motion interference blood oxygen saturation algorithm for the microcontroller based on AC and DC analysis, named de-trended FFT.
Methods:
An experiment was conducted to compare the de-trended FFT algorithm with two other algorithms commonly used in the time and frequency domains. In the experiment, participants' oxygen saturation levels were calculated from Photoplethysmography (PPG) signals that were recorded continuously. Meantime, five types of hand motions were conducted, including hand trembling movements, horizontal hand movements, vertical hand movements, finger tapping, and finger bending, with each state lasting 2 minutes.
Results:
Results show significant performance of de-trended FFT in SpO2 calculation (P < 0.05), in both accuracy and stability.
Conclusion:
De-trended FFT stands out in both mean deviation and variance by eliminating trending influence when compared with the other two algorithms. The motion interference's influence on SpO2 calculation mainly comes from the AC component, not the DC.
More Related Videos
09:42Author Spotlight: Using Motor Imagery Brain-Computer Interface to Improve Motor and Cognitive Function in Stroke Patients
Published on: September 1, 2023
08:35Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.