Multiprobe devices for exercise transcutaneous oxymetry in patients complaining claudication: interest and limits of
Samir Henni1, Christophe Colas-Ribas, Isabelle Signolet
1Laboratory for Vascular Investigations, Mitovasc Institut, University Hospital of Angers, Angers, France - piabraham@chu-angers.fr.
Background:
Exercise transcutaneous oximetry (PtcO2) can argue for a vascular origin of exercise-related pain in atypical unusual localizations, such as lumbar, thigh or foot pain. We used five probes as a standard for treadmill PtcO2 tests. Recent commercially available devices now include up to eight probes. We aimed at analyzing the potential interest of positioning a 6th probe on the area of unusual localization during exercise PtcO2 tests.
Methods:
We retrospectively analyzed our data with a 6-probes device, using as a standard: one probe on the chest, one on each buttock and one on each calf. The sixth probe was positioned either: in the lumbar median position (N.=342), on the anterior lower part of the thigh (N.=391) or on the dorsum of the foot (N.=155) in patients complaining exercise-induced pain including the back, thigh or foot respectively. Results on the sixth probe at the limb were compared to result of the standard adjacent probe. A positive test (abnormal result) was defined as a minimal value of the DROP-index lower than minus 15 mmHg.
Results:
Prevalence of positive results on the 6th probe with negative results on the adjacent standard probe was 2.3% at the lumbar site, 3.8% at the thigh and 12.3% at the foot level.
Conclusions:
Atypical localizations are rarely but sometimes associated to isolated positive exercise PtcO2 results and may justify the use of more than five probes in some patients, specifically for foot pain.
More Related Videos
04:44Author Spotlight: Enhancing Vascular Function and Physical Capacity in Cardiovascular Disease Through Novel Interventions and NIRS Technology
Published on: March 22, 2024
04:20Integration of Brain Tissue Saturation Monitoring in Cardiopulmonary Exercise Testing in Patients with Heart Failure
Published on: October 1, 2019
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Assessing Blood pressure in the Leg
Preparation:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...