Sensitivity of a helical diode array dosimeter to Volumetric Modulated Arc Therapy delivery errors
Sankar Arumugam1, Aitang Xing1, Tony Young1
1Liverpool and Macarthur Cancer Therapy Centres and Ingham Institute, Sydney, NSW, Australia.
Summary
The ArcCHECK dosimeter effectively detects delivery errors in Volumetric Modulated Arc Therapy (VMAT). It can identify Multi Leaf Collimator (MLC) position and dose errors above 3mm and 3% respectively, ensuring treatment accuracy.
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- Volumetric Modulated Arc Therapy (VMAT) is a complex radiotherapy technique requiring precise delivery.
- Ensuring accuracy in VMAT delivery is crucial for patient safety and treatment efficacy.
- Dosimetric verification tools are essential for quality assurance in VMAT.
Purpose of the Study:
- To evaluate the sensitivity of the ArcCHECK dosimeter in detecting various delivery errors during VMAT.
- To assess the capability of ArcCHECK to identify Multi Leaf Collimator (MLC) position and dose delivery inaccuracies.
Main Methods:
- Simulated gantry-independent, gantry-dependent, and control point-specific MLC and dose errors in prostate and head and neck VMAT plans.
- Utilized ArcCHECK for measurements and analyzed local and global gamma pass rates with 2%/2mm and 3%/3mm tolerances.
- Calculated the clinical impact of the simulated errors.
Main Results:
- ArcCHECK detected minimum 3mm MLC errors and 3% output errors for gantry-independent scenarios with 2%/2mm tolerance.
- Minimum 3mm MLC errors and 3% dose errors were identifiable for gantry-dependent scenarios.
- Errors of 10mm MLC shift and 100% dose error at 4 control points were detected.
Conclusions:
- ArcCHECK, with 2%/2mm criteria and either local or global gamma analysis, reliably detects VMAT delivery errors exceeding specified thresholds.
- The findings support the confident clinical use of ArcCHECK for VMAT quality assurance.
- This study validates ArcCHECK's role in identifying clinically relevant VMAT delivery deviations.