Multiple breath washout cannot be used for tidal breath parameter analysis in infants
Georgia L Banton1, Graham L Hall1, Mark Tan1
1Telethon Kids Institute, University of Western Australia, Subiaco, Australia.
Pediatric Pulmonology
|October 6, 2015
Summary
Multiple breath washout (MBW) testing with SF6 gas mixture alters infant tidal breathing parameters. A separate room air trace is necessary before MBW to ensure accurate analysis of ventilation distribution in infants.
Area of Science:
- Pediatric Pulmonology
- Respiratory Physiology
Background:
- Multiple breath washout (MBW) using sulfur hexafluoride (SF6) is standard for infant ventilation assessment.
- The impact of SF6 on infant tidal breathing parameters during MBW is not well understood.
- Investigating these changes is crucial for accurate respiratory assessments in infants.
Purpose of the Study:
- To determine if SF6 gas mixture alters tidal breathing parameters in infants during MBW testing.
- To assess the necessity of a separate tidal breathing measurement in room air prior to MBW.
Main Methods:
- Comparison of tidal breathing in room air versus during MBW with SF6 in healthy, preterm, and cystic fibrosis infants.
- Analysis of key respiratory parameters: inspiratory/expiratory times, time to peak flows, tidal volume, respiratory rate, and minute ventilation.
Main Results:
- Significant increases in breath times (inspiratory and expiratory) were observed in healthy and cystic fibrosis infants during MBW wash-in.
- Decreased respiratory rate and minute ventilation were noted in healthy and cystic fibrosis infants during MBW wash-in.
- Preterm infants showed reduced minute ventilation during both wash-in and wash-out phases of MBW.
Conclusions:
- SF6 gas mixture significantly alters tidal breathing parameters in infants during MBW testing.
- Measuring a separate tidal breathing trace in room air before MBW is essential.
- This pre-MBW measurement ensures the rigor and accuracy of derived tidal breath indices for clinical analysis.