Related Experiment Video
Updated: Apr 1, 2026

A Millimeter Scale Flexural Testing System for Measuring the Mechanical Properties of Marine Sponge Spicules
Published on: October 11, 2017
Effect of bending direction on the mechanical behaviour of 3.5 mm String-of-Pearls and Limited Contact Dynamic
J Benamou, R M Demianiuk, S Rutherford
1Loïc M. Déjardin, DVM, MS, Diplomate ACVS & ECVS, College of Veterinary Medicine, Michigan State University, 736 Wilson Road, East Lansing, MI 48824, United States,
Objective:
To compare the bending properties of String-of-Pearls® (SOP) and Limited Contact Dynamic Compression Plate® (LC-DCP) constructs in orthogonal bending directions.
Methods:
3.5 mm SOP and LC-DCP plates were fixed to a bone model simulating a comminuted tibial fracture. Specimens were non-destructively tested in both mediolateral and craniocaudal bending for 10 cycles. Bending stiffness and total angular deformation were compared using parametric analyses (p <0.05).
Results:
For both constructs, stiffness was significantly less when bending moments were applied against the thickness of the plates (mediolateral bending) than against the width (craniocaudal bending). When compared to the mediolateral plane, bending constructs in the craniocaudal plane resulted in a 49% (SOP group) and 370% (LC-DCP group) increase in stiffness (p <0.001). Mediolateral bending stiffness was significantly greater in the SOP than the LC-DCP constructs. Conversely, in craniocaudal bending, SOP constructs stiffness was significantly less than that of the LC-DCP constructs. The differences between the two constructs in total angular deformation had an identical pattern of significance.
Clinical Significance:
This study found that SOP showed less variability between the orthogonal bending directions than LC-DCP in a comminuted fracture model, and also described the bi-planar bending behaviour of both constructs. Although not exhibiting identical bending properties in both planes, SOP constructs had a more homogenous bending behaviour in orthogonal loading directions. The difference between the SOP with a circular cross sectional shape compared to the rectangular shape of standard plates is probably responsible for this difference.
More Related Videos
04:20Author Spotlight: Enhancing Accuracy and Reproducibility in Whole Bone Bending Tests
Published on: September 1, 2023
07:15A Testing Platform for Durability Studies of Polymers and Fiber-reinforced Polymer Composites under Concurrent Hygrothermo-mechanical Stimuli
Published on: December 11, 2014
Related Concept Videos
Bending
In pure bending, the bending stress in a beam is calculated based on the bending moment and...
Members Made of Elastoplastic Material
As the bending moment...
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each material's...
Plastic Behavior
Deformations in a Transverse Cross Section
As the material stretches, it expands or contracts in orthogonal directions to the load. This phenomenon varies...