Related Experiment Video
Updated: Apr 1, 2026

Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Accuracy of Tympanic Temperature Measurement in Firefighters Completing a Simulated Structural Firefighting Task
Toby Keene1, Matt Brearley2, Beth Bowen1
11Australian Capital Territory Ambulance Service,Quality, Safety, and Risk Unit,Canberra,Australia.
Introduction:
In the course of their duties, firefighters risk heat stroke and other medical conditions due to exertion in high-temperature environments. Infrared tympanic temperature measurement (TTym) is often used by Emergency Medical Services (EMS) to assess the core body temperature of firefighters. The accuracy of TTym in this setting has been called into question. Hypothesis/Problem This study aimed to examine the accuracy of TTym for core body temperature assessment at emergency firefighting events compared with gastrointestinal temperature measurement (TGI) as measured by ingestible thermometers.
Methods:
Forty-five (42 male, three female) professional urban firefighters from an Australian fire service completed two 20-minute work periods in a 100°C (± 5°C) heat chamber while wearing personal protective clothing (PPC) and breathing apparatus (weighing approximately 22 kg). Measurements were taken immediately before entering, and on exiting, the heat chamber. Tympanic temperature was assessed by an infrared tympanic thermometer and TGI was measured by ingestible sensor and radio receiver.
Results:
Complete data were available for 37 participants. Participant temperatures were higher on exiting the heat chamber than at baseline (TTym: 35.9°C (SD=0.7) vs 37.5°C (SD=0.8); TGI: 37.2°C (SD=0.4) vs 38.6°C (SD=0.5)). Tympanic temperature underestimated TGI on average by 1.3°C (SD=0.5) before entering the chamber and by 1.0°C (SD=0.8) following the exercise. Using pooled data, the average underestimation was 1.2°C (SD=0.7).
Conclusion:
Tympanic thermometers cause an unreliable measure of core body temperature for firefighters engaged in fire suppression activities. Accurate and practical measures of core body temperature are required urgently.
More Related Videos
04:34Mouse Body Temperature Measurement Using Infrared Thermometer During Passive Systemic Anaphylaxis and Food Allergy Evaluation
Published on: September 14, 2018
04:54Using a Combination of Indirect Calorimetry, Infrared Thermography, and Blood Glucose Levels to Measure Brown Adipose Tissue Thermogenesis in Humans
Published on: June 2, 2023
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...