Efficacy of Different Techniques for Removing Debris from Endodontic Files Prior to Sterilization
Saeid Nosouhian1, Farshad Bajoghli2, Mahmoud Sabouhi1
1Assistant Professor, Department of Prosthodontics, Dental Implant Research Center, School of Dentistry, Isfahan University of Medical Sciences, Isfahan, Iran.
Background:
Dental instruments, like endodontic files, are hardly disinfected. The aim of this study was to evaluate the efficacy of mechanical, chemical, and ultrasonic (in combinations or separately) techniques for removing debris from files, prior to sterilization.
Materials And Methods:
Totally, 90 new endodontic files with size of 15, 25, and 40 (30 files of each one) were sterilized and one files of each one kept as negative control (NC) group. The rest of files (29 files of each size) were divided into five groups after endodontic therapies and prior to autoclaving: (Positive control [PC]: Without interventions, A: Cleaned with scouring sponge soaked in chlorhexidine 0.2%, B: Stored in Micro 10 enzyme, C: Subjected to Micro 10 enzyme in both conventional and ultrasonic way for 15 min, D: Decontaminated by Micro 10 enzyme with ultrasonic). Finally, the samples were observed under a metallographic microscope, and the data were analyzed by Tuckey, paired t-test, two-way ANOVA tests using SPSS software version 15 at a significant level of 0.05.
Results:
Significant differences were observed in heads of the files among groups C and PC (P = 0.02), and high amount of debris were seen in the shafts of groups A and D (P < 0.001). The amount of remaining debris were significant in the shafts of sizes 15 (P < 0.001) and 25 (P = 0.01).
Conclusion:
Using Micro 10 in both ultrasonic and conventional methods were acceptable for removing debris from the files. Furthermore, higher amounts of debris were found in the shafts and heads of files with lower sizes (15 and 25).
More Related Videos
03:07Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
03:10Pilot In Vitro Study to Assess Cleaning Ability and Effects of Different Decontamination Methods on Implant Surfaces
Published on: November 21, 2025
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Key Techniques in Microbiology
Hand hygiene
Hand washing...