Influence on Visual Quality of Intraoperative Orientation of Asymmetric Intraocular Lenses
Purpose:
To evaluate visual quality when changing the intraocular orientation of the Lentis Mplus LS-312MF nonrotational symmetric +3.00 diopters aspheric multifocal intraocular lens ([IOL] Oculentis GmbH, Berlin, Germany) in normal eyes.
Methods:
An artificial eye was used to measure the in vitro wavefront of the IOL. The corneal topography of 20 healthy patients was obtained. For each eye, a computational analysis simulated the implantation of the IOL. The modulation transfer function (MTF) and an image quality parameter (visually modulated transfer function [VSMTF] metric) were calculated for a 5.0-mm pupil and for three conditions: distance, intermediate, and near vision. The procedure was repeated for each eye after a rotation of the IOL with respect to the cornea from 0° to 360° in 1° steps.
Results:
Statistical analysis showed significant differences in mean VSMTF values between orientations for distance vision. Optimal orientation of the IOL (different for each eye) showed a mean improvement of 58% ± 19% (range: 20% to 121%) in VSMTF values with respect to the worst possible orientation. For these orientations, intermediate and near vision quality were statistically indistinguishable. The MTFs were different between orientations, showing a mean difference of approximately 5 cycles per degree in the maximum spatial frequencies that can be transferred between the best and the worst orientations for distance vision.
Conclusions:
The results suggest that implantation of this nonrotational symmetric IOL should improve visual outcomes if it is oriented to coincide with a customized meridian. A simple, practical method is proposed to find an approximation to the angle that an Mplus IOL should be inserted.
More Related Videos
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Related Concept Videos
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview
Angle Closure Glaucoma: Treatment
Focusing of Light in the Eye