Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
Nineteen-Millimeter Bioprosthetic Aortic Valves Are Safe and Effective for Elderly Patients With Aortic Stenosis
Zain Khalpey1, Patrick O Myers2, Siobhan McGurk3
1Division of Cardiac Surgery, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts; Division of Cardiothoracic Surgery, Department of Surgery, University of Arizona College of Medicine, Tucson, Arizona.
Background:
Replacing a stenotic aortic valve with 19-mm bioprostheses remains controversial owing to potential patient-prosthesis mismatch concerns. We report a single-center 10 year experience with 19-mm bioprosthetic valves implanted in elderly patients. We hypothesized patients would have acceptable in-hospital and long-term outcomes.
Methods:
Between January 2002 and December 2011, 257 patients underwent aortic valve replacement with a 19-mm prosthesis, of whom 182 had available follow-up echocardiographic studies. Mean age was 77.4 ± 8.4 years, and 10 of 257 (4%) were male. Outcomes of interest included early and late mortality, peak and mean aortic valve gradients, and left ventricular mass regression.
Results:
Operative mortality was 3.5% (9 of 257). Median postoperative echocardiographic time was 16 months. On follow-up echocardiography, mean peak aortic valve gradient decreased from 76 ± 27 mm Hg preoperatively to 32 ± 13 mm Hg and the mean gradient decreased from 46 ± 17 mm Hg to 18 ± 8 mm Hg (both p < 0.001) Mean left ventricular mass decreased from 191 g to 162 g (p < 0.001). Postoperative survival did not differ significantly between patients who met the criteria for patient-prosthesis mismatch and those who did not (p = 0.607).
Conclusions:
In a series of elderly patients with aortic stenosis who were implanted with 19-mm bioprosthetic valves, long-term follow-up showed significant left ventricular mass regression and peak and mean aortic valve gradient reductions. The use of 19-mm aortic valves is safe and efficacious for elderly patients with a small aortic root.