A Randomized Controlled Trial Comparison of NeuroSENSE and Bispectral Brain Monitors During Propofol-Based Versus
Julie Bresson1, Etienne Gayat, Gracee Agrawal
1From the *Department of Anesthesiology, Hopital Foch, Suresnes, France; †Department of Anesthesiology, Hopital Foch, University Versailles Saint-Quentin en Yvelines, France; ‡Department of Anesthesiology and Intensive Care, Hopital Lariboisiere, Paris, France; §Clinical Epidemiology and Biostatistics U171, INSERM, Paris, France; ‖Department of Anesthesiology, Hopital Lariboisiere, University Paris Diderot, Paris, France; ¶NeuroWave Systems Inc., Cleveland Heights, Ohio; #Outcomes Research Consortium, Cleveland, Ohio; and **Department of Electroencephalography, Hopital Foch, Suresnes, France.
Background:
NeuroSENSE is a depth of anesthesia monitor that uses automated electroencephalogram quantification. The Wavelet-based Anesthetic Value for Central Nervous System (WAVCNS) index calculated by this monitor is based on wavelet analysis of a normalized electroencephalogram signal in the γ-frequency band. The aim of this study was to determine the extent of disagreement between the Bispectral Index (BIS) and the WAVCNS index during propofol-based and sevoflurane-based maintenance of general anesthesia in a routine surgical population.
Methods:
Patients undergoing elective surgery were enrolled in the study and randomly assigned to receive either propofol or sevoflurane for the maintenance of anesthesia and remifentanil in both groups. Anesthesiologists were blinded to monitors in both groups. Discordance between the 2 monitors was assessed by the count of discrepancy in recommendation (DR) (type 1 defined as one parameter <40 and the other >60, or type 2 defined as BIS and WAVCNS values on different sides of a threshold [40 or 60]) and also by the proportion of agreement (P0) between WAVCNS and BIS, obtained every 5 seconds, in 3 categories of index (<40, 40-60, and >60).
Results:
The analyzed data set consisted of 22 patients (36,872 data pairs) in the propofol group and 24 patients (32,826 data pairs) in the sevoflurane group. The type 1 DR rarely occurred in both the groups (<1%); however, the median (interquartile range) type 2 DR was significantly more frequent in the propofol group (20.6% [7.0-36.9] vs 4.5% [2.3-12.4]; P = 0.0005). The median difference in P0 was 11.53% (95% confidence interval, 0.57-21.32). Major disagreement between WAVCNS index and BIS was related to the weight of burst suppression pattern for the index calculation.
Conclusions:
Disagreement between BIS and NeuroSENSE during the maintenance of general anesthesia was worse in the propofol group than that in the sevoflurane groups. The disagreement increases during deep anesthesia or in the occurrence of burst suppression.
More Related Videos
07:54Drug-Induced Sleep Endoscopy DISE with Target Controlled Infusion TCI and Bispectral Analysis in Obstructive Sleep Apnea
Published on: December 6, 2016
08:49Assessing Changes in Volatile General Anesthetic Sensitivity of Mice after Local or Systemic Pharmacological Intervention
Published on: October 16, 2013
Related Concept Videos
General Anesthesia: Overview
General anesthesia induces unconsciousness in the whole body, while the others target specific areas or sensations. It is administered to minimize adverse effects, maintain...
Stages of General Anesthesia