Continuous remote monitoring of postoperative vital signs with the Biobeat patch: A multicenter prospective
Inès Kouidri1, Cécile Landais2, Audrey Solis3
1Department of Anesthesiology and Pain Medicine, Hôpital Foch, Suresnes, France.
Objective:
This study, conducted under standard clinical conditions, evaluated the Biobeat wearable sensor in postoperative patients monitored over multiple days without human intervention. The study aimed to assess technical performance (vital sign data collection and artifact frequency), reliability (agreement with nurse-measured values), the sensor's ability to detect vital sign abnormalities, patient satisfaction, and incidence of adverse skin reactions.
Design:
This multicenter prospective observational study included patients who underwent abdominal surgery with a planned postoperative ward stay between December 2020 and December 2022.
Methods:
Alongside routine care-comprising (2-3 times daily nurse assessments of blood pressure (BP), pulse rate (PR), respiratory rate (RR), peripheral oxygen saturation (SpO2), and temperature)-participants wore a precordial Biobeat sensor. The sensor transmitted data wirelessly to a cloud-based repository via Wi-Fi without human intervention.
Results:
Of 109 enrolled patients, 90 were included in the analysis. The median recording duration was 63.8 hours (interquartile range: 42.6-72.3 hours). Data loss occurred in 45% of the recording time, and pulse oximetry data were absent in 16% of the available measurements. Artifacts were infrequent, comprising 277 of 35,998 measurement points. Compared with nurse measurements, Biobeat showed small mean differences for most variables, except for RR and temperature, yet exhibited wide limits of agreement across all variables. Clarke Error Grid analysis revealed excellent concordance between Biobeat and nurse measurements for SpO2 and PR, with lower agreement for mean BP, RR, and temperature. The sensor's continuous monitoring detected more vital sign abnormalities, including severe hypotension (mean BP < 60 mmHg), in 10.0% of patients compared to 3.3% with nurse measurements (P = 0.04). The patients reported high satisfaction and no adverse skin reactions were observed.
Conclusions:
Significant data loss presents a substantial challenge for analysis, underscoring the critical need for improved data transmission and flow control measures in study protocols and clinical deployment.
Registry:
ClinicalTrials.gov (NCT04585178, October 14, 2020).
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:01A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Holter Monitor: 24-Hour Monitoring
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Peripheral Artery Disease V: Postoperative Nursing Management
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: