Arterial input functions (AIFs) measured directly from arteries with low and standard doses of contrast agent, and
Shiyang Wang1, Xiaobing Fan1, Milica Medved1
1Department of Radiology, University of Chicago, Chicago, IL 60637.
Abstract:
Measurements of arterial input function (AIF) can have large systematic errors at standard contrast agent doses in dynamic contrast enhanced MRI (DCE-MRI). We compared measured AIFs from low dose (AIFLD) and standard dose (AIFSD) contrast agent injections, as well as the AIF derived from a muscle reference tissue and artery (AIFref). Twenty-two prostate cancer patients underwent DCE-MRI. Data were acquired on a 3T scanner using an mDixon sequence. Gadobenate dimeglumine was injected twice, at doses of 0.015 and 0.085 mmol/kg. Directly measured AIFs were fitted with empirical mathematical models (EMMs) and compared to the AIF derived from a muscle reference tissue (AIFref). EMMs accurately fitted the AIFs. The 1st and 2nd pass peaks were visualized in AIFLD, but not in AIFSD, thus the peak and shape of AIFSD could not be accurately measured directly. The average scaling factor between AIFSD and AIFLD in the washout phase was only 56% of the contrast dose ratio (~6:1). The shape and magnitude of AIFref closely approximated that of AIFLD after empirically determined dose-dependent normalization. This suggests that AIFref may be a good approximation of the local AIF.
More Related Videos
09:43In vivo Near Infrared Fluorescence NIRF Intravascular Molecular Imaging of Inflammatory Plaque, a Multimodal Approach to Imaging of Atherosclerosis
Published on: August 4, 2011
10:02Quantification of Atherosclerotic Plaque Activity and Vascular Inflammation using [18-F] Fluorodeoxyglucose Positron Emission Tomography/Computed Tomography FDG-PET/CT
Published on: May 2, 2012