Parallel Plating of Simulated Distal Humerus Fractures Demonstrates Increased Stiffness Relative to Orthogonal
Phillip A Taylor1, John R Owen, Charles P Benfield
1Departments of Orthopaedic Surgery and Biomedical Engineering, Orthopaedic Research Laboratory, Virginia Commonwealth University, Richmond, VA.
Objectives:
This study compared the stiffness of precontoured parallel and orthogonal locking plate configurations in cyclic torsion and bending, and then extension to failure.
Methods:
Tests were conducted on 9 matched pairs of cadaveric humeri. A 10 mm block of bone was excised from the distal humerus metaphysis to simulate comminution, and fractures were repaired in matched fashion using parallel or orthogonal Biomet ALPS distal humerus locking plates (Biomet, Inc, Warsaw, IN). Specimens were tested at 0.5 degree per second up to ±2 N-m in internal and external rotation for 20 cycles, then in flexion and extension cantilevered bending at 1 mm/s to ±50 N, followed by bending at 1 mm/s to failure in extension.
Results:
Torsional stiffness of the parallel configuration group was greater than the orthogonal configuration in both internal and external rotation (P < 0.0001). Also, stiffness in bending was significantly greater in the parallel configuration group in both flexion and extension (P < 0.0001). In extension to failure testing, the parallel plate construct stiffness was significantly greater than the orthogonal configuration (P < 0.005).
Conclusions:
The parallel plate configuration demonstrated significantly greater stiffness than the orthogonal plate configuration in torsion and bending using locked distal humerus plates. This greater stiffness may prove desirable in the postoperative management of patients with comminuted distal humerus fractures, providing a stable anatomic reconstruction of the joint to allow early range of motion.
More Related Videos
07:35Creating Rigidly Stabilized Fractures for Assessing Intramembranous Ossification, Distraction Osteogenesis, or Healing of Critical Sized Defects
Published on: April 11, 2012
06:38Establishment of a Segmental Femoral Critical-size Defect Model in Mice Stabilized by Plate Osteosynthesis
Published on: October 12, 2016
Related Concept Videos
Bones of the Upper Limb: Humerus
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...