Evaluation of plasma-based transmit coils for magnetic resonance
1C.J. Gorter Center for High Field MRI, Leiden University Medical Center, The Netherlands.
Abstract:
In this work a new concept for designing transmit coils for magnetic resonance using a plasma is introduced. Unlike conventional coils, a plasma can be turned on and off, eliminating electrical interactions between coils, and enabling radiofrequency-invisibility when desired. A surfatron has been designed to produce a surface-mode wave which propagates along the inner surface of a commercial fluorescent lighting tube. NMR spectra and images have been produced using the plasma as the transmit coil and a copper-based monopole to receive the signal. The transmit efficiency of the plasma tube was estimated, and is currently much lower than that of an equivalently-sized metal-based structure: however, there are many potential methods for increasing the efficiency using a custom-built plasma tube.