The 10m incremental shuttle walk test is a highly reliable field exercise test for patients referred to cardiac
Lisa C Hanson1, Nicholas F Taylor2, Helen McBurney3
1La Trobe Rural Health School, La Trobe University, Bendigo, Victoria, Australia.
Objectives:
To determine the retest reliability of the 10m incremental shuttle walk test (ISWT) in a mixed cardiac rehabilitation population.
Design:
Participants completed two 10m ISWTs in a single session in a repeated measures study. Ten participants completed a third 10m ISWT as part of a pilot study.
Setting:
Hospital physiotherapy department.
Participants:
62 adults aged a mean of 68 years (SD 10) referred to a cardiac rehabilitation program.
Main Outcome Measures:
Retest reliability of the 10m ISWT expressed as relative reliability and measurement error. Relative reliability was expressed in a ratio in the form of an intraclass correlation coefficient (ICC) and measurement error in the form of the standard error of measurement (SEM) and 95% confidence intervals for the group and individual.
Results:
There was a high level of relative reliability over the two walks with an ICC of .99. The SEMagreement was 17m, and a change of at least 23m for the group and 54m for the individual would be required to be 95% confident of exceeding measurement error.
Conclusions:
The 10m ISWT demonstrated good retest reliability and is sufficiently reliable to be applied in practice in this population without the use of a practice test.