Importance of Pulse Repetition Frequency Adjustment for 3- and 4-Dimensional Power Doppler Quantification
Andre H Miyague1, Theo Z Pavan1, Carlos A Soares1
1Department of Obstetrics and Gynecology, Medical School of Ribeirão Preto (A.H.M., C.A.S., C.O.N., W.P.M.), and Department of Physics, School of Philosophy, Sciences, and Letters of Ribeirão Preto (T.Z.P.), University of São Paulo, Ribeirao Preto, Brazil; Department of Obstetrics and Gynecology, University Hospital Evangelico, Curitiba, Brazil (A.H.M.); Woman and Fetal Medicine Institute, Curitiba, Brazil (A.H.M.); Ultrasonography and Retraining Medical School of Ribeirão Preto, Ribeirão Preto, Brazil (C.A.S., C.O.N., W.P.M.); Department of Obstetrics and Gynecology, University Hospitals Leuven, Leuven, Belgium (L.D.C.); School of Women's and Children's Health, University of New South Wales, Randwick, New South Wales, Australia (A.W.W.); and Department of Maternal-Fetal Medicine, Royal Hospital for Women, Randwick, New South Wales, Australia (A.W.W.).
Objectives:
To determine the influence of the pulse repetition frequency (PRF) and wall motion filter on the 3-dimensional (3D) power Doppler vascularization-flow index (VFI) and volumetric pulsatility index (PI) obtained from spatiotemporal image correlation (STIC) data sets acquired from a common carotid artery of a healthy participant.
Methods:
We acquired 11 STIC data sets, 1 for each PRF value ranging from 0.6 to 9.0 kHz. Vascularization-flow index and volumetric PI values were determined from the 440 static 3D data sets contained in these STIC data sets. Additionally, 3 sets of radio-frequency data were acquired for offline processing of different wall motion filter values for PRF values of 0.6, 3.3, and 10 kHz.
Results:
We constructed VFI curves and observed 2 patterns: a flattened pattern with a low PRF and a triphasic pattern with a high PRF, correlating with the known pulsed wave Doppler profile of this vessel. Volumetric PI values were around 0 for low PRF settings and increased with increasing PRF. Analysis of the radiofrequency data showed that increasing wall motion filter values gradually filtered out the low-velocity power Doppler signals while retaining the higher-velocity ones, allowing the distinction of integrated power Doppler signal velocity throughout the cardiac cycle.
Conclusions:
We conclude that the PRF and wall motion filter dramatically influence 3D power Doppler indices and the volumetric PI, and the use of PRF values in which minimum VFI values are measured during the diastolic phase in the spectral Doppler wave may validate the use of the volumetric PI.
Related Concept Videos
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Doppler Effect - II
Doppler Effect - I