Related Experiment Video
Updated: Mar 30, 2026

Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Adverse events in contemporary continuous-flow left ventricular assist devices: A multi-institutional comparison
John M Stulak1, Mary E Davis2, Nicholas Haglund2
1Division of Cardiovascular Surgery, Mayo Clinic College of Medicine, Rochester, Minn.
Objectives:
We review differences in the incidence and timing of adverse events in patients implanted with continuous-flow left ventricular assist devices (LVADs), using the Mechanical Circulatory Support Research Network registry.
Methods:
From May 2004 to September 2014, a total of 734 patients (591 men; median age: 59 years) underwent primary continuous-flow LVAD implantation at our institutions. Patients implanted with the HeartMate II (HMII) (560 [76%] patients), compared with the HeartWare ventricular assist device (HVAD; 174 [24%]) were more often receiving destination therapy (47% vs 20%; P < .01), had a lower preoperative creatinine level (1.2 vs 1.3; P = .01), and had less median preoperative right ventricular dysfunction (mild vs moderate; P < .01). Ischemic etiology, prior sternotomy, and median INTERMACS profile were similar.
Results:
Overall mortality was 54 of 734 (7.4%); 41 of 560 (7.3%) in the HMII group, and 13 of 174 (7.5%) in the HVAD group (P = .95). Follow-up was available in 100% of early survivors for a median of 1 year (max: 10 years) and a total of 1120 patient-years of support (HMII: 940 patient-years [median: 1.1 years, max: 5.3 years] and HVAD: 180 patient-years [median: 0.6 year, max: 10.4 years]). On multivariable analysis, GI bleeding (P = .63), any infection (P = .32), driveline infection (P = .10), and pump thrombus (P = .64) were similar between devices while HeartWare HVAD was associated with higher risk of stroke (HR: 1.8, [1.25, 2.5], P = .003).
Conclusions:
In this pooled analysis, a trend was found for higher incidence of percutaneous driveline infections in patients treated with the HMII; a higher incidence of stroke and time-related cumulative risk of any infection and stroke was found in patients treated with the HVAD, which was independently associated with higher stroke risk.
More Related Videos
05:18Implantation of Left Ventricular Assist Device (LVAD) in Juvenile Landrace Swine: A LVAD Implantation Model of Pediatric Heart Failure
Published on: January 16, 2026
07:39Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Heart Failure VI: Adjunct Therapies
Cardiomyopathy II: Dilated Cardiomyopathy
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Continuous Renal Replacement Therapy
Heart Failure V: Medical Management