Related Experiment Video
Updated: Jun 24, 2026

A Novel Rescue Technique for Difficult Intubation and Difficult Ventilation
Published on: January 17, 2011
Impending Complete Airway Obstruction from a Reinforced Orotracheal Tube: a Case Report
Nima Hosseinzadeh1, Shahram Samadi2, Mihan Jafari Javid2
1School of Medicine, Tehran University of Medical Sciences, Tehran, Iran.
Abstract:
Reinforced tubes are commonly used to minimize the opportunity of upper airway obstruction in patients at risk. There are a few reports of the airway obstruction resulted from kinked reinforced tubes. This report describes the obstruction of a reinforced tube in an adult patient who underwent tonsillectomy. Under general anesthesia; an armoured endotracheal tube was inserted into the trachea uneventfully. A few minutes after starting the surgery, the anesthesia machine detected a high airway pressure and an increased ETCO2 (end-tidal CO2) up to 50 mmHg. Further evaluation showed spiral wire damage resulted from Mouth Gag device that led to airway obstruction. Early anticipation of the complications leads to proper management of such critical and life threatening complications and prevention of hypoxia, hypercapnia, pneumothorax, and pulmonary edema. Based on our experience using an armoured endotracheal tube in tonsillectomy does not guarantee a safe airway and intensive monitoring is necessary.
Related Concept Videos
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of the...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques