A physician's office-based digoxin assay (Seralyzer) evaluated for interference by endogenous digoxin-like
1Department of Pathology, Jewish Hospital of St. Louis, Washington University Medical Center, MO 63110.
Annals of Clinical and Laboratory Science
|May 1, 1989
Summary
The Ames Seralyzer digoxin test shows minimal interference from digoxin-like immunoreactive factors (DLIF) in most patients. This suggests the Seralyzer is a reliable method for measuring digoxin levels, even in populations with high DLIF concentrations.
Area of Science:
- Clinical Chemistry
- Immunoassay Technology
- Pharmacokinetics
Background:
- Digoxin is a critical medication for heart conditions, requiring precise monitoring.
- Digoxin-like immunoreactive factors (DLIF) can interfere with digoxin testing, leading to inaccurate results.
- Physician's office laboratory (POL) analyzers need to be evaluated for accuracy and interference.
Purpose of the Study:
- To evaluate the Ames Seralyzer digoxin immunoassay for interference by DLIF.
- To compare the Seralyzer's performance with a known DLIF-detecting assay (Abbott TDx).
- To assess the impact of DLIF on digoxin measurements in various patient populations.
Main Methods:
- Sera from normal, renal, hepatic, pregnant, and neonatal patients were analyzed.
- Samples were tested using the Ames Seralyzer and Abbott TDx digoxin immunoassays.
- Interference was quantified by measuring apparent digoxin values and DLIF detection potency.
Main Results:
- Only 4.7% of non-digoxin patients showed apparent digoxin >0.2 ng/mL on the Seralyzer, versus 75% on the TDx.
- The TDx assay was 17-fold more potent in detecting DLIF from adrenal extracts compared to the Seralyzer.
- Digoxin presence reduced DLIF interference potency, particularly for the TDx assay.
Conclusions:
- The Ames Seralyzer demonstrates significantly less DLIF interference compared to the Abbott TDx.
- The Seralyzer provides reliable digoxin measurements, even in patients with potential DLIF.
- Reduced DLIF cross-reactivity in the presence of digoxin explains comparable results between assays for patients on digoxin.