Cerebral CT Venography Using a 320-MDCT Scanner With a Time-Density Curve Technique and Low Volume of Contrast Agent:
Shubhabrata Biswas1, Arun Chandran1, Sue Roughley1
11 All authors: Department of Neuroradiology, The Walton Centre for Neurology and Neurosurgery, Lower Ln, Fazakerley, Liverpool, L9 7LJ, United Kingdom.
Objective:
The purpose of this study was to compare a cerebral CT venography (CTV) technique performed on a 320-MDCT scanner with the use of a time-density curve (TDC) and a small volume of IV contrast medium (35 mL, with 15 mL used for the test bolus) with CTV performed using an established reference technique in which scanning is performed at a fixed time delay of 30 seconds with the use of a larger volume of contrast medium (100 mL).
Materials And Methods:
The time of peak enhancement was determined from the TDC generated from a scan in which a test bolus dose was used. CTV was performed at the time of peak enhancement. The diagnostic quality of 31 CTV venograms acquired using this technique was compared retrospectively with the diagnostic quality of 29 CTV venograms obtained at a fixed time delay of 30 seconds. The densities in the major venous sinuses and the degree of arterial contamination on the scans acquired using the two techniques were compared using objective and semiobjective methods. The semiobjective assessments were made independently by two neuroradiologists.
Results:
Attenuation was higher in the venous structures seen on CTV images acquired using the TDC technique. Of the scans obtained using the TDC technique, the proportion deemed to be of good quality, on the basis of a grading scale, was statistically significantly higher (p < 0.05). Also, the degree of arterial contamination was statistically significantly lower (p < 0.05). The interrater agreement for semiobjective assessments ranged from good to very good.
Conclusion:
We describe a CTV technique performed using a low volume of IV contrast medium and a TDC on a 320-MDCT scanner. This technique provides better venous opacification and lower arterial contamination compared with use of the fixed time-delay technique.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation I: X-ray and CT
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...