Related Experiment Video
Updated: Mar 29, 2026

10:02
Submillisecond Conformational Changes in Proteins Resolved by Photothermal Beam Deflection
Published on: February 18, 2014
9.5K
Photon beam quality correction factors for the NE2571A and NE2581A thimble ionization chambers using PENELOPE
Fabián Erazo1, Antonio M Lallena2
1Departamento de Física Atómica, Molecular y Nuclear, Universidad de Granada, E-18071 Granada, Spain; Instituto del Cáncer - SOLCA, Cuenca, Ecuador.
Abstract:
The beam quality correction factor kQ,Q0 and the perturbation factor pQ for photon beams were calculated for the NE2571A and NE2581A ionization chambers, using the Monte Carlo simulation code PENELOPE. Results are compared to those quoted for the NE2571 and NE2581 chambers in previous works. Both kQ,Q0 and pQ obtained for NE2571A and NE2581A chambers agree with those of their predecessors NE2571 and NE2581 ones.