Hepatocellular Carcinoma Screening With Computed Tomography Using the Arterial Enhancement Fraction With
Adrian Thomas Huber1, Frederik Schuster, Lukas Ebner
1From the *Department of Diagnostic, Interventional and Pediatric Radiology, University Hospital Inselspital Bern, Bern, Switzerland; †Department of Polyvalent and Oncologic Radiology, University Hospital Pitié-Salpêtrière, Paris, France; ‡Department of Radiology, Duke University Medical Center, Durham, North Carolina; and §Institute of Pathology, University of Bern, Bern, Switzerland.
Objective:
The aim of this study was to investigate the performance of the arterial enhancement fraction (AEF) in multiphasic computed tomography (CT) acquisitions to detect hepatocellular carcinoma (HCC) in liver transplant recipients in correlation with the pathologic analysis of the corresponding liver explants.
Materials And Methods:
Fifty-five transplant recipients were analyzed: 35 patients with 108 histologically proven HCC lesions and 20 patients with end-stage liver disease without HCC. Six radiologists looked at the triphasic CT acquisitions with the AEF maps in a first readout. For the second readout without the AEF maps, 3 radiologists analyzed triphasic CT acquisitions (group 1), whereas the other 3 readers had 4 contrast acquisitions available (group 2). A jackknife free-response reader receiver operating characteristic analysis was used to compare the readout performance of the readers. Receiver operating characteristic analysis was used to determine the optimal cutoff value of the AEF.
Results:
The figure of merit (θ = 0.6935) for the conventional triphasic readout was significantly inferior compared with the triphasic readout with additional use of the AEF (θ = 0.7478, P < 0.0001) in group 1. There was no significant difference between the fourphasic conventional readout (θ = 0.7569) and the triphasic readout (θ = 0.7615, P = 0.7541) with the AEF in group 2. Without the AEF, HCC lesions were detected with a sensitivity of 30.7% (95% confidence interval [CI], 25.5%-36.4%) and a specificity of 97.1% (96.0%-98.0%) by group 1 looking at 3 CT acquisition phases and with a sensitivity of 42.1% (36.2%-48.1%) and a specificity of 97.5% (96.4%-98.3%) in group 2 looking at 4 CT acquisition phases. Using the AEF maps, both groups looking at the same 3 acquisition phases, the sensitivity was 47.7% (95% CI, 41.9%-53.5%) with a specificity of 97.4% (96.4%-98.3%) in group 1 and 49.8% (95% CI, 43.9%-55.8%)/97.6% (96.6%-98.4%) in group 2. The optimal cutoff for the AEF was 50%.
Conclusion:
The AEF is a helpful tool to screen for HCC with CT. The use of the AEF maps may significantly improve HCC detection, which allows omitting the fourth CT acquisition phase and thus making a 25% reduction of radiation dose possible.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography