Advances in Patient Self-Monitoring of Blood Glucose
1Pfützner Science & Health Institute, Germany andreas.pfuetzner@pfuetzner-mainz.com.
Abstract:
In 2 articles of the present issue, Bendini et al report about performance results obtained with 2 blood glucose monitoring systems of the Contour Next platform. Using several analysis methods, the authors demonstrate a very high accuracy, which meets all actual regulatory performance criteria. With consistent MARD results < 5% under daily routine conditions, this meter platform is finally fulfilling the accuracy request as set forth by the American Diabetes Association already in the late 1980s. This meter platform is representative for the successful effort of the device manufacturers who were consequently improving the analytical performance of blood glucose meters during the Past 2 decades, starting with an MARD of 12-15% at the end of the past century and reaching an excellent accuracy < 5% today.