Related Experiment Video
Updated: Mar 29, 2026

Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Short and long term outcomes of 200 patients supported by continuous-flow left ventricular assist devices
Athanasios Tsiouris1, Gaetano Paone1, Hassan W Nemeh1
1Athanasios Tsiouris, Jeffrey A Morgan, Section of Cardiac Surgery, Yale University School of Medicine, New Haven, CT 06520, United States.
Aim:
To study the institutional experience over 8 years with 200 continuous-flow (CF) - left ventricular assist devices (LVAD).
Methods:
We evaluated our institution's LVAD database and analyzed all patients who received a CF LVAD as a bridge to transplant (BTT) or destination therapy from March 2006 until June 2014. We identified 200 patients, of which 179 were implanted with a HeartMate II device (Thoratec Corp., Pleasanton, CA) and 21 received a Heartware HVAD (HeartWare Inc., Framingham, MA).
Results:
The mean age of our LVAD recipients was 59.3 years (range 17-81), 76% (152/200) were males, and 49% were implanted for the indication of BTT. The survival rate for our LVAD patients at 30 d, 6 mo, 12 mo, 2 years, 3 years, and 4 years was 94%, 86%, 78%, 71%, 62% and 45% respectively. The mean duration of LVAD support was 581 d (range 2-2595 d). Gastrointestinal bleeding (was the most common adverse event (43/200, 21%), followed by right ventricular failure (38/200, 19%), stroke (31/200, 15%), re exploration for bleeding (31/200, 15%), ventilator dependent respiratory failure (19/200, 9%) and pneumonia (15/200, 7%). Our driveline infection rate was 7%. Pump thrombosis occurred in 6% of patients. Device exchanged was needed in 6% of patients. On multivariate analysis, preoperative liver dysfunction, ventilator dependent respiratory failure, tracheostomy and right ventricular failure requiring right ventricular assist device support were significant predictors of post LVAD survival.
Conclusion:
Short and long term survival for patients on LVAD support are excellent, although outcomes still remain inferior compared to heart transplantation. The incidence of driveline infections, pump thrombosis and pump exchange have declined significantly in recent years.
More Related Videos
Related Concept Videos
Cardiomyopathy V: Interprofessional Care
Heart Failure VI: Adjunct Therapies
Cardiomyopathy II: Dilated Cardiomyopathy
Heart Failure V: Medical Management
Pathophysiology of Heart Failure