Efficacy of Electronic Foramen Locators in Controlling Root Canal Working Length during Rotary Instrumentation
Lorena Arruda Parente1, Martin D Levin2, Rodrigo Ricci Vivan3
1School of Dental Medicine of Sobral, UFC - Universidade Federal do Ceará, Sobral, CE, Brazil.
Brazilian Dental Journal
|December 10, 2015
Summary
Electronic foramen locators (EFLs) precisely control root canal working length during rotary preparation. However, precision decreased when penetration was limited, with the Root ZX II (RZX) AAR failing in all instances.
Area of Science:
- Endodontics
- Dental Materials Science
Background:
- Accurate control of root canal working length is crucial for successful endodontic treatment.
- Electronic foramen locators (EFLs) are widely used to determine and maintain working length during instrumentation.
- Variations in EFL performance based on device and preparation level require further investigation.
Purpose of the Study:
- To evaluate the efficacy of two EFLs, Root ZX II (RZX) and Propex II (PRO), in controlling root canal working length during rotary instrumentation.
- To assess the reliability of these devices at different preparation levels (0.0 mm and -1.0 mm from the apical foramen).
Main Methods:
- Forty-eight human mandibular bicuspids were prepared using the Protaper rotary system (crown-down technique).
- EFLs (RZX in auto-reverse mode, PRO with MPAS-10R) monitored preparation at 0.0 mm and -1.0 mm levels.
- Post-preparation, file tip to apical foramen distance was measured after sectioning the apical third.
Main Results:
- At 0.0 mm, both RZX and PRO achieved 100% precision.
- At -1.0 mm, PRO showed 66.7% precision, while RZX showed 0.0% precision.
- RZX demonstrated significantly poorer results (0.96±0.11 mm) compared to PRO (0.43±0.23 mm) at -1.0 mm.
Conclusions:
- EFLs are precise in maintaining root canal working length when reaching the apical foramen.
- Both devices exhibited decreased precision when penetration was limited to -1.0 mm.
- The RZX AAR mode failed to maintain working length in all tested instances at -1.0 mm, highlighting potential reliability issues in specific scenarios.