Objective:
To establish a rapid determination method with gas chromatography-mass spectrometer (GC-MS) for thiodiglycolic acid (TDGA), a vinyl chloride (VCM) biomarker.
Methods:
A high- sensitivity determination method was established using a moderate methyl esterification instead of methyl esterification of highly toxic diazo reaction.
Results:
The standard curve regression linear equation of the method was: y=8460.5x-4758.2, the linear coefficient was 0.999 7, the minimum quantity concentration was 2.0 µg/L, the range of precision value was 0.81%-2.38%, and the average recovery of standard addition was 99.0%-102.9%.
Conclusion:
This method reduces the risk of traditional methyl esterification, improves the determination sensitivity compared with the GC-FPD method, and meets the determination requirement of TDGA.