[Simultaneous determination of three organic fluorides in workplace air by portable gas chromatography-mass
Objective:
To establish a method for rapid determination of organic fluorides in the air of a fluorine chemical plant using portable gas chromatography-mass spectrometer (GC-MS).
Methods:
Standard samples of monochlorodifluoromethane, tetrafluoroethylene, and hexafluoropropylene of different concentrations were prepared by static volumetric method with high-purity nitrogen as the diluent gas. The samples were injected into the GC-MS by a hand-held probe. Retention time and characteristic ion were used for qualitative analysis, and the area of selected ion peak was used for quantitative analysis. The standard curves were then created for quantitative determination of the three organic fluorides.
Results:
The linear ranges for monochlorodifluoromethane, tetrafluoroethylene, and hexafluoropropylene by the method were 0.39-7.72, 0.45-8.84, and 0.61-12.20 mg/m3, respectively, the average recovery rates for the three concentrations were 102.8%, 96.0%, and 106.5%, respectively, and the average deviations were 2.1%, 5.1%, and 2.4%, respectively.
Conclusion:
The portable GC-MS can be used for the simultaneous qualitative and quantitative analysis of monochlorodifluoromethane, tetrafluoroethylene, and hexafluoropropylene in the workplace air, and the method is simple, fast, and accurate.
More Related Videos
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Flame Photometry: Lab
High-Performance Liquid Chromatography: Types of Detectors