Some factors that can lead to poor peak shape in hydrophilic interaction chromatography, and possibilities for their
James C Heaton1, David V McCalley1
1Centre for Research in Biosciences, University of the West of England, Frenchay, Bristol BS16 1QY, UK.
Abstract:
Some factors which present difficulties for obtaining good peak shape in hydrophilic interaction chromatography (HILIC) were studied. The effect of injection solvent composition and volume was systematically investigated using a selection of weak and stronger basic compounds on a hybrid bare silica phase. Increasing the mismatch between the injection solvent (range 95-0% ACNv/v) and the mobile phase (maintained at 95% ACNv/v) gave increasing deterioration in peak shape. With the 2.1mm ID columns used, injections in the mobile phase of increasing volume (1-20 μL) gave poorer peak shape, but the magnitude of the effect was considerably smaller than that of solvent mismatch over this range. Some solute structural features such as galloyl (trihydroxy benzene), catechol (benzene diol) and phosphate (in nucleotides) gave serious peak tailing, attributed to interactions with metals in the stationary phase or the chromatographic hardware. These undesirable effects can be moderated by including complexing agents in the mobile phase, by changing the stationary phase chemistry, or by altering the mobile phase pH.
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Chromatographic Methods: Terminology