Related Experiment Video
Updated: Mar 28, 2026

Closure of a Patent Foramen Ovale PFO: An Intervention Sequence
Published on: December 23, 2022
Early experience with the Occlutech PLD occluder for mitral paravalvar leak closure through a hybrid transapical
Radwa Bedair1, Gareth J Morgan, Vinayak Bapat
1University Hospitals Bristol NHS Foundation Trust, Bristol, United Kingdom.
Aims:
We sought to evaluate the feasibility and efficacy of hybrid transapical closure of paravalvar mitral leaks using a new Occlutech PLD occluder in patients with heart failure and/or haemolytic anaemia.
Methods And Results:
Retrospective analysis of clinical and procedural data was undertaken for patients who had attempted closure of paravalvar mitral leaks via a hybrid transapical approach with the Occlutech PLD occluder. Eight patients (four males, median age 69 years) underwent closure of 10 mitral paravalvar leaks using eight Occlutech PLD occluders and two AMPLATZER Vascular Plugs (AVP II). Successful deployment, with significant reduction of the paravalvar leak was achieved in seven patients with short procedure (median 131 min) and fluoroscopy times (median 22 min). One patient had mechanical interference with prosthetic valve function, requiring surgery. Another patient with a high EuroSCORE (48.8%) died of multi-organ failure two days after the procedure. Clinical improvement in either heart failure or haemolysis was seen in all discharged patients.
Conclusions:
In our series of patients with challenging anatomy, the Occlutech PLD occluders performed well when implanted via a hybrid transapical approach. Further work is needed to assess this methodology fully for a wider population and to assess other deployment approaches for this promising new occluder.
Related Concept Videos
Mitral Regurgitation I: Introduction
Mitral Valve Prolapse I: Introduction
Mitral Stenosis III: Medical Management
Mitral Stenosis I: Introduction
Mitral Valve Prolapse II: Assessment and Management
Mitral Regurgitation III: Medical Management